CAUTION

• If a large pulling force is applied to the input/output signal wires connected to the

MX100 and MW100, the terminal or signal wire may break. To prevent this from

happening, secure all the wiring cables to the installation panel.

• Do not apply a voltage exceeding the value indicated below to the input

terminals of the universal input modules. Doing so can damage the modules.

• Maximum input voltage

Voltage range of 1 VDC or less, TC, RTD, and DI (contact): ± 10 VDC

Voltage range of 2 VDC or more, and DI (LEVEL): ± 120 VDC

• Maximum common mode voltage

Input to ground:

600 VACrms (50/60 Hz)

Between channels:

250 VACrms (50/60 Hz) (-H04)

• Maximum common mode noise voltage

Between channels:

120 VACrms (-M10, “b” terminals are common to

all channels. They are connected together internally.)

• Do not apply a voltage exceeding the value indicated below to the input

terminals of the DCV/TC/DI input module. Doing so can damage the module.

• Maximum input voltage

Voltage range of 1 VDC or less, TC, and DI (contact): ± 10 VDC

Voltage range of 2 VDC or more, and DI (LEVEL): ± 120 VDC

• Maximum common mode voltage

Input to ground:

600 VACrms (50/60 Hz)

• Maximum common mode noise voltage

Between channels:

120 VACrms

• Do not apply a voltage exceeding the value indicated below to the input

terminals of the Four-wire RTD resistance input module. Doing so can damage

the module.

• Maximum input voltage

Voltage range of 1 VDC or less, RTD, resistance, and DI (contact): ± 10 VDC

Voltage range of 2 VDC or more, and DI (LEVEL): ± 120 VDC

• Maximum common mode voltage

Input to ground:

600 VACrms (50/60 Hz)

• Maximum common mode noise voltage

Between channels, DC voltage, DI: 120 VACrms

Between channels, RTD, resistance: 50 VACrms

• Wiring of the Strain Input Module (-NDI)

When connecting a bridge head, in order that the empty weight of the cable

does not exceed 5 kg, ensure that the cable does not hang down more than 1.5

m (the distance to the floor). If the cable hangs longer than 1.5 m, secure the

cable to the installation panel or some other location.

• Do not apply a voltage exceeding the value indicated below to the input

terminals of the strain input modules. Doing so can damage the modules.

• Maximum input voltage: ± 10 VDC

• Maximum common mode voltage

Between channels: 30 VACrms (50/60 Hz)

Input to ground: 250 VACrms (-B12, -B35), 30 VACrms (-NDI) (50/60 Hz)

• Do not apply a voltage exceeding the value indicated below to the input

terminals of the pulse input module. Doing so can damage the module.

• Maximum input voltage: ± 10 VDC

• Maximum common mode voltage

Input to ground: 250 VACrms (50/60 Hz)

• When using the pulse input module with contact input, the measured signal

becomes easily affected by wiring impedance at high speed. The cable should

be approximately 25 m or less when the pulse width is 0.05 ms, or 500 m or less

at 0.5 ms.

The wiring impedance varies depending on the such things as the cable length,

type, and wiring conditions.