3 error messages and corrective actions, Error, Message – Yokogawa DAQWORX User Manual

Page 25: 3 error messages and corrective actions -5, Error -5, Message -5, 3 error messages and correctve actons, Index

Advertising
background image

3-5

IM WX1-08E

1

2

3

Index

Deta

ls of Funct

ons

3.3

Error Messages and Correctve Actons

A message (such as an error message) may appear during operation. The following

describes the meanings of the messages and their corrective actions.

Error

Code

Message

Correctve Actons

E211

Cannot write to the file.

Confirm that the disc capacity and file system are correct.

E212

Cannot read file.

Confirm that a file exists, that the software supports it, and whether the file system
is correct.

E213

Cannot open the file.

Confirm that a file exists, that the software supports it, and whether the file system
is correct.

E401

Communication error.

Check whether the power to the connected measuring instruments with which
you wish to open communications is ON, and whether the cables are connected
correctly. Also check the following.

• For Ethernet

Is the address correct, is the TCP/IP protocol installed in Windows, and is the

Ethernet card installed correctly? Is the address correct, is the TCP/IP protocol
installed in Windows, and is the Ethernet card installed correctly?

• For RS-232 and RS-422A

Are the baud rate, port numbers (COM1–COM9), and addresses (RS422)

correct? Is the PC serial port valid? Is the cable type correct?

E402

Communication timeout.

E403

Communication open error.

Check whether the power to the connected measuring instruments with which
you wish to open communications is ON, and whether the cables are connected
correctly. Also check the following.

• For Ethernet

Is the address correct, is the TCP/IP protocol installed in Windows, and is the

Ethernet card installed correctly?

• For RS-232 and RS-422A

Are the baud rate, port numbers (COM1–COM9), and addresses (RS422)

correct? Is the PC serial port valid? Is the cable type correct?

E501

Invalid license number.

With the Gate series, the number of licenses purchased is the limit for the number
of programs run at the same time.

E1001

Login failed.

Check the user name and password on the µR10000.

E1002

Invalid password.

Check the user name and password on the µR10000.

E1003

Invalid password.

Check the user name and password on the µR10000.

E1010

A user with the same name is
already logged in.

Check the user name and password on the µR10000.

Message

Code

Message

M1201

Model determination completed successfully.

M1210

Save the setting changes and try again.

Advertising
This manual is related to the following products: