Page 4-13 “Clock Display”

The date and time can be displayed on sub-display 2.
According to the set time in 3.10, “Setting the Date and Time” (see page 3-35), the current date and
time is displayed.
The display shows the month, day, year and hour, minute, second in this sequence.

Page 13-5 “Error Codes”

The following error codes have been added to the list.

Error Codes

Error

Corrective Action

012

None of the channels specified in execution of

initial balance were strain input channels.

Enter a correct channel.

013

Attempted “REP RECALL START” without hourly,

daily, or monthly report data.

032

Set to contiguous channels on the power monitor

module.

Enter a correct channel.

047

Entered a wiring method for which there is no

setting for the power monitor module.

Enter the correct wiring method.

107

Attempted to change ranges or time while report

was starting.

Do not make changes.

131

Media write error

Exchange the medium.

137

Attempted to start computation or execute a

procedure during saving of settings to a medium

or while reading from a medium.

Start the procedure or computation after

completion of the media operation.

138

Media drive error

Exchange the medium. If the error occurs again

after exchanging the medium, servicing is

required.

146

An initial balancing error occurred during

calibration.

Check the wiring and try calibration again.

Page 13-8 “calibrations”

For “Strain measurement” under “Connection,” the wiring diagram for the bridge box used
for the DU500-14 has been changed to the following:

Connect CH2 only

Switch the

bridge head

R2

R4

R1

R3

ON

1 2 3 4 5

1

2

3

4

5

6

7

8

OFF
SW

Bridge head

(701955 or 701956)

SW1

OFF

SW2

OFF

SW3

OFF

SW4

ON

SW5

OFF

Page 14-3 “Standard Computation Functions”

Scaling

Measurement accuracy for scaling: measurement accuracy for scaling (digits) = measurement
accuracy (digits) × scaling span (digits) / measurement span (digits) + 2 digits. Numbers below
the decimal point are rounded up.

±(((0.05/100) Ч 5000) + 2) Ч (2000/4000) + 2 = ±4.25
Measurement accuracy = ±5 digits = ±0.005 V