Page 14-8 “normal operation conditions, Page 14-9 “emc eonformity standards, Page 14-20 “measuring ranges and accuracies – Yokogawa DR240 User Manual

Page 5: Page 14-21 “specifications of strain input module

Advertising
background image

IM DR232-01E-9A 5/5

for IM DR232-01E 9th Edition

Page 14-8 “Normal Operation Conditions”

Installation category based on IEC61010-1, CSA22.2 No.61010-1

II*

1

Pollution degree based on IEC61010-1, CSA22.2 No.61010-1

2*

2

Warm-up time

At least 30 minutes after power switch-on.

*

1

Describes a number which defines a transient overvoltage condition. It implies the regulation for impulse

withstand voltage. “II” applies to electrical equipment which is supplied from fixed installations like distribution
boards.

*

2

Describes the degree to which a solid, liquid, or gas which deteriorates dielectric strength or surface

resistivity is adhering. “2” applies to normal indoor atmosphere. Normally, only non-conductive pollution
occurs.

Page 14-9 “EMC Eonformity Standards”

Please refer to these specifications instead of the one printed in the user's manual.

Safety and EMC Standards

CSA

CSA22.2 No.61010-1, installation category II, pollution degree 2

UL

UL61010-1 (CSA NRTL/C)

C-Tick

EN55011 compliance, Class A, Group 1

KC marking

Electromagnetic wave interference prevention standard, electromagnetic wave protection

standard compliance

Page 14-12 “Measurement range, accuracy and resolution”

Input

Measurement (digital display)

DC Voltage

20mV

60mV

200mV

2V

:

–20.000 to 20.000mV

–60.00 to 60.00mV

–200.00 to 200.00mV

–2.0000 to 2.0000V

:

Type

Measurement accuracy

Maximum

resolution

(0.05% of rdg + 5digits)

(0.05% of rdg + 2digits)

(0.05% of rdg + 2digits)

(0.05% of rdg + 2digits)

:

Measurement range

1µV

10µV

10µV

100µV

:

Page 14-15, 14-19, 14-21, 14-23, 14-25 “Specifications of Module”

Installation Category (Overvoltage Category)

CAT II (IEC61010-1, CSA22.2 No.61010-1)

Measurement Category

CAT II (IEC61010-1, CSA22.2 No.61010-1)

Page 14-20 “Measuring Ranges and Accuracies”

Gauge Method Measurement Range Type Rated Measurement Range

Accuracy

Resolution

Single-gauge

2000µε

-2000 to 2000µε

0.5% of Range

0.1µε

20000µε

-20000 to 20000µε

0.3% of Range

1µε

200000µε

-200000 to 200000µε

0.3% of Range

10µε

Two-gauge

1000µε

-1000 to 1000µε

0.5% of Range

0.1µε

10000µε

-10000 to 10000µε

0.3% of Range

1µε

100000µε

-100000 to 100000µε

0.3% of Range

10µε

Four-gauge

500µε

-500 to 500µε

0.5% of Range

0.1µε

5000µε

-5000 to 5000µε

0.3% of Range

1µε

50000µε

-50000 to 50000µε

0.3% of Range

10µε

Page 14-21 “Specifications of Strain Input Module”

Accesorry
Bridge head: 701955, 701956

Advertising
This manual is related to the following products: