SAFETY AND EMC STANDARDS

CAUTION

This instrument is for Measurement Category I (CAT.I).

Do not use it for measurements in locations falling

under Measurement Categories II, III, and IV.

Internal Wiring

Outlet

Entrance

Cable

III

T

I

II

IV

Measurement

category

Description

Remarks

I

CAT.I For measurements performed

on circuits not directly connected

to MAINS.

II

CAT.II For measurements performed

on circuits directly connected to

the low-voltage installation.

Appliances,

portable

equipments,

etc.

III

CAT.III For measurements performed in

the building installation.

Distribution

board, circuit

breaker, etc.

IV

CAT.IV For measurements performed

at the source of the low-voltage

installation.

Overhead

wire, cable

systems, etc.

An analog input signal is measurement category I (CAT.I).

Rated transient overvoltage: 1500 V

(Note)

Note

This is a reference safety standard value for

Measurement Category I of IEC/EN/CSA/UL61010-1.

This value is not necessarily a guarantee of instrument

performance.

EMC standards: Complies with EN61326.

The above conformed instrument is only for voltage of 15 to 30

V DC.

* The instrument continues to operate at a measurement

accuracy of within ±20% of the range during testing.