Maintenance, Safety and emc standards, 1 calibration apparatus – Yokogawa JUXTA VJS7 User Manual

Page 5: 2 calibration procedure, Caution

Advertising
background image

5

IM 77J01S07-01E 5th Edition

Mar. 02, 2012-00

8. MAINTENANCE

The product starts running immediately when the power is turned

on; however, it needs 10 to 15 minutes of warm-up before it

meets the specified performance.

8.1 Calibration Apparatus

● 6-dial decade resistance boxes × 2 units

(Yokogawa Meters & Instruments 279301 or equivalent)

● A digital multimeter (DMM) (Yokogawa 7561 or equivalent)

● A calibrator (Yokogawa Meters & Instruments CA150 or

equivalent)

● A precision resistor of 250 Ω ±0.01%, 1W

● Setting tool for adjustment (Refer to

“4.2 Connector for

Communication” in this manual.)

8.2 Calibration Procedure

1. Connect the instruments as shown below. First adjust the

output-1 signal and then the output-2 signal.

10

11

3 2 1

4

5

6

7

8

9

Power supply

11

10

8

L

+

N

GND

DMM

2

5

Output-2

R

+

Calibrator

2

5

Alarm-1

ALM1

COM

+

Calibrator

6

5

Alarm-2

ALM2

COM

+

DMM

7

9

Output-1

R

+

R: For current output using 250 Ω

precision resistor

1

4

3

Input

6 dial

variable

resistor
6 dial

variable

resistor

100%

R1

R1

R2

R2

0%

2. Operate the variable resistor according to the specifications

(total resistance, 0%, 100% resistance) of the potentiometer

to be connected. Keeping total resistance constant, change

each value of R1 and R2 to apply the resistance equivalent to

0, 25, 50, 75, and 100% of the input range to the converter.

3. Check to see the corresponding output voltages are 0, 25, 50,

75, and 100% respectively and within the specified accuracy

rating. (“R” is used for current output.)

For alarm output, check the relay action by the alarm indicator

lamp or resistance of output terminals.

● Use the setting tool (VJ77 Parameter Setting Tool or JHT200

Handy Terminal) to adjust the input/output signals.

User’s Manual for VJ77 [Document No.: IM 77J01J77-01E];

however, use the VJ77 of version R1.04 or later.

User’s Manual for JHT200 [Document No.: IM JF81-02E]

SAFETY AND EMC STANDARDS

CAUTION

This instrument is for Measurement Category I (CAT.I).

Do not use it for measurements in locations falling

under Measurement Categories II, III, and IV.

Internal Wiring

Outlet

Entrance

Cable

III

T

I

II

IV

Measurement

category

Description

Remarks

I

CAT.I For measurements performed

on circuits not directly connected

to MAINS.

II

CAT.II For measurements performed

on circuits directly connected to

the low-voltage installation.

Appliances,

portable

equipments,

etc.

III

CAT.III For measurements performed in

the building installation.

Distribution

board, circuit

breaker, etc.

IV

CAT.IV For measurements performed

at the source of the low-voltage

installation.

Overhead

wire, cable

systems, etc.

An analog input signal is measurement category I (CAT.I).

Rated transient overvoltage: 1500 V

(Note)

Note

This is a reference safety standard value for

Measurement Category I of IEC/EN/CSA/UL61010-1.

This value is not necessarily a guarantee of instrument

performance.

EMC standards: Complies with EN61326.

The above conformed instrument is only for voltage of 15 to 30

V DC.

* The instrument continues to operate at a measurement

accuracy of within ±20% of the range during testing.

Advertising