IM 77J04R31-01E 1st Edition : 2006.08.31-00

9.2

Instrument for Adjustments

• 6-dial variable resistor (Yokogawa Meters & Instruments’ 2793 01 or equivalent): 1

9.3

Input Adjustment

9.3.1

Connecting the Adjustment Instrument

Connect the input and supply voltage as shown in the figure below.

10 11

1

2

3

9

8

7

6

5

4

7

8

N–

L+

(SUPPLY)

4

6

5

B

B

A

Input signal

Power supply

6-dial

variable

resistor

9.3.2

Adjusting Inputs

Carry out input adjustments between two points, or the minimum value (BL: input adjustment
point LOW) and maximum value (BH: input adjustment point HIGH) that have been set
within the adjustment range (see Section 9.1). This subsection describes an example of
making an input adjustment between two points within the range of -50.0 to 150.0

°

C with the

MVRK range code No. set to “62” (instrument input range: -199.9 to 200.0

°

C).

Press to display

"LOC."

Press to display

"-1."

The MVRK enters the operable status as soon as the power is turned on, but requires 10 to 15 minutes
of warm-up to meet the performance requirements.

When you press the SET/ENT key for more than 3 sec.
with the Operation Parameter Screen displayed, the Setup
Parameter Screen 1 appears.

When the power is turned on, the PV screen of the
Operation Parameter Screen appears.

Press the SET/ENT key to display parameter "BL."
In this case, the DATA display shows "-199.9," the
minimum value of instrument input range of the set range
code No.

Press the UP key to display "-50.0" on the DATA display.

Press the SET/ENT key to accept this data.

Press the SET/ENT key to display the next parameter.

This causes parameter "AL" and the measured value to
appear alternately.

Press the UP or DOWN key.
The decimal point blinks.

Press the SET/ENT key.

The value before adjustment (-52.2) appears on the DATA
display.

Press the SET/ENT key to display the next parameter "BH."

Press this key for
more than 3 sec.

The decimal point blinks.

The decimal point blinks during
data change.

The decimal point blinks during
data change.

Press the SET/ENT key to display parameter "LOC."

Press the DOWN key to display "-1" on the DATA display.

Press the SET/ENT key to display the Setup Parameter
Screen 2.

Setup Parameter Screen 1

Setup Parameter Screen 2

Operation Parameter Screen

Power ON

Using the 6-dial variable resistor, apply resistance equivalent

to the input -50

°

C to the MVRK.

In this case, the DATA display shows "200.0," the
maximum value of instrument input range of the set range
code No.

Press the DOWN key to display "150.0" on the DATA
display.

Press the SET/ENT key to accept this data.

Press the SET/ENT key to display the next parameter.

The decimal point blinks during
data change.

Press .

Press to

display "BL."

Press to display

"-50.0."

Press .

Press .

The measured value and
parameter AL are
displayed alternately.

Press or .

Press .

Press .

"-52.2" appears on the
DATA display.

Press to display

"150.0."

Press .

Press .

The previously adjusted value
will be reset.

The decimal point blinks during
data change.

This causes parameter "AH" and the measured value to
appear alternately.

Press the UP or DOWN key.
The decimal point blinks.

Press the SET/ENT key.

The value before adjustment (153.9) appears on the DATA
display.

This completes the input adjustment for the MVRK.

Press the SET/ENT key to display parameter "RST."

Press the UP key to display "ON" on the DATA display.

Press the SET/ENT key.
This resets the previously adjusted value.

Using the 6-dial variable resistor, apply resistance equivalent

to the input 150

°

C to the MVRK.

If additional re-adjustment must be made, take the following steps and then perform the
procedure above.

Press to display

"RST."

Press to display

"ON."

Press .

This completes the input
adjustment.

The decimal point blinks.

The measured value and
parameter AH are displayed
alternately.

Press or .

Press .

10. SETTING ACTIVE COLOR PV DISPLAY (PV DISPLAY COLOR CHANGING FUNCTION)

10.1 Setting Ranges and Factory-Set Values

Setup Parameter Screen 1

When PV display color mode (PCM) is 6 or 7:
PCL+1digit to 9999
When PV display color mode (PCM) is 8 or 9:
-100.0 to 100.0% of the measured input range
* The setting range depends on the setting of the

decimal point position (SDP).

Parameter Name

Setting Range

Factory-Set

Value

Parameter

Symbol

Ϫ

Ϫ

High limit for PV
display color change
(PCH)

When PV display color mode (PCM) is 6 or 7:
-1999 to PCH-1digit
When PV display color mode (PCM) is 8 or 9:
-100.0 to 100.0% of the measured input range
* The setting range depends on the setting of the

decimal point position (SDP).

Low limit for PV
display color change
(PCL)

Setup Parameter Screen 2

Parameter Name

Setting Range

Factory-Set

Value

Parameter

Symbol

1

PV display color mode
(PCM)

*: PV display color is changed linking to the setting range of high limit (PCH) and low limit (PCL) for PV display color change.

0: Fixed in green
1: Fixed in red
2: Link to alarm 1
(under normal condition: green; at alarm status: red)
3: Link to alarm 1
(under normal condition: red; at alarm status: green)
4: Link to alarm 1 and alarm 2
(under normal condition: green; at alarm status: red)
5: Link to alarm 1 and alarm 2
(under normal condition: red; at alarm status: green)
6: PV limit*
(when more than PCL, less than PCH: green;
when PCL or less, PCH or more: red)
7: PV limit*
(when more than PCL, less than PCH:red;
when PCL or less, PCH or more: green)
8: SP deviation*
(when more than SP-PCL, less than SP+PCH: green;
when SP-PCL or less, SP+PCH or more: red)
9: SP deviation*
(when more than SP-PCL, less than SP+PCH: red;
when SP-PCL or less, SP+PCH or more: green)
10: Link to alarm 1 to alarm 4
(under normal condition: green; at alarm status: red)
11: Link to alarm 1 to alarm 4
(under normal condition: red; at alarm status: green)

10.2 Setting Active Color PV Display

This section describes an example of setting the PV display color mode (PCM) to “6,”
high limit for PV display color change (PCH) to “70.0” and low limit (PCL) to “20.0.”
The procedure below begins with the condition in which the Setup Parameter Screen 2 is
displayed.

When you press the SET/ENT key for more than 3 sec.
with the Operation Parameter Screen displayed, the Setup
Parameter Screen 1 appears.

Press the SET/ENT key to display parameter "PCH."

Press the UP or DOWN key to display "70.0" on the DATA
display.

Press this key for
more than 3 sec.

The decimal point blinks during
data change.

Operation Parameter Screen

This completes the process for
setting the high limit for PV display
color change.

Press to display

"PCH."

Press or

to display "70.0."

Press .

Press to display
"PCL."

Press the SET/ENT key to accept the high limit for PV
display color change "70.0."

Press the SET/ENT key to display parameter "PCL."

Press the UP or DOWN key to display "20.0" on the DATA
display.

The decimal point blinks during
data change.

This completes the process for
setting the low limit for PV display
color change.

Press or

to display "20.0."

Press .

Press for more
than 3 sec.

Press the SET/ENT key to accept the low limit for PV
display color change "20.0."

Press the SET/ENT key for more than 3 sec.
This causes the Operation Parameter Screen to appear.

Press the SET/ENT key to display parameter "PCM."

Press the UP or DOWN key to display "6" on the DATA
display.

The decimal point blinks during
data change.

This completes the process for
setting the PV display color mode.

Press to display

"PCM."

Press or

to display "6."
"""

Press .

Press for more
than 3 sec.

Press the SET/ENT key to accept PV display color
mode "6."

Press the SET/ENT key for more than 3 sec.
This causes the Operation Parameter Screen to appear.

Setting the PV display color mode

Setting the high and low limits for PV display color change

Setup Parameter Screen 1

Setup Parameter Screen 2

To the Operation Parameter Screen