Yokogawa JUXTA FT5V Thermocouple Converter (Free Range Type) User Manual

User's manual, Checking product specifications and package, General

Advertising
background image

User's

Manual



Thank you for purchasing the JUXTA Signal Conditioner.

Please read through this manual before use for correct handling.

Network Solutions Business Division

2-9-32, Naka-cho Musashino-shi, Tokyo 180-8750 Japan

Phone: +81-422-52-7179 Facsimile: +81-422-52-6619

IM 77J08T05-01E

1st Edition Jan. 2007 (YK)

CAUTIONARY NOTES FOR SAFE USE OF

THE PRODUCT

This User’s Manual should be carefully read before installing and

operating the product. The following symbol is used on the product

and in this manual to ensure safe usage.

This symbol is displayed on the product when it is
necessary to refer to the User's Manual for information
on personal and instrument safety. This symbol is
displayed in the User's Manual to indicate precautions
to avoid danger to the operator, such as an electric
shock.

The following symbols are used only in this manual.

NOTE

D r a w s a t t e n t i o n t o e s s e n t i a l i n f o r m a t i o n f o r
understanding
the operations and/or functions of the product.

CHECKING PRODUCT SPECIFICATIONS

AND PACKAGE

(1) Checking the Model and Product Specifications

Check that the model and specifications indicated on the
nameplate attached to the main unit are as ordered.

(2) Packaged Items

Check that the package contains the following items:

FT5A or FT5V: 1
Mounting block: 2
Tag number and range label : 1 each
Mounting screw (M4 x 0.7): 2
User’s Manual (this manual : IM 77J08T05-01E)

GENERAL

The FT5A/FT5V is a free range type thermocouple converter that
is connected to an IEC/JIS-standard thermocouple (TC), such as
a Type K, T, E, J, R, S, B or N thermocouple to convert tempera-
ture signals into isolated
DC current or DC voltage signals.

Selection of input type (sensor type), temperature unit (°C, K), I/
O range setting, burnout setting, output adjustment, RJC on/off
and I/O monitoring can be made using the optional Parameter
Setting Tool (VJ77) or Handy Terminal (JHT200).





MODEL AND SUFFIX CODES

Model

Suffix codes

Description

FT5A

-  *C / Thermocouple converter

(Free Range Type)

FT5V
Input

-1

K (CA)

-2

T (CC)

-3

E (CRC)

-4

J (IC)

-5

R

-6

S

-7

B (RH)

-8

N

-0

Custom order

Output

A

4 to 20 mA DC

FT5A

B

2 to 10 mA DC

C

1 to 5 mA DC

D

0 to 20 mA DC

E

0 to 16 mA DC

F

0 to 10 mA DC

G

0 to 1mA DC

Z

Custom order

1

0 to 10 mV DC

FT5V

2

0 to 100 mV DC

3

0 to 1 V DC

4

0 to 10 V DC

5

0 to 5 V DC

6

1 to 5 V DC

7

-10 to 10 V DC

0

Custom order

Style code

*C

Style C

Burnout

/BU Up
/BD Down
/BN None

Temperature unit

/DF

Fahrenheit (°F) display function

Models FT5A, FT5V

Thermocouple Converter

(Free Range Type)

Advertising
This manual is related to the following products: