Condec DLR334 User Manual

Page 16

Advertising
background image

PAGE 16

74845

! TEMPERATURE CORRECTIONS DATA ENTRY AND RECALL:

COMMANDS:

D1E{}/D1R and D2E{}/D2R

DATA FORMAT (1):

<zr>|<zl1>|<zl2>|<zl3>|<zh1>|<zh2>|<zh3>|<sr>|<sl1>|<sl2>|<sl3>|

<sh1>|<sh2>|<sh3>

DATA FORMAT (2):

<tr>|<tl1>|<tl2>|<tl3>|<th1>|<th2>|<th3>|<fzl1>|<fzl2>|<fzl3>|<fsl1>|

<fsl2>|<fsl3>|<fzh1>|<fzh2>|<fzh3>|<fsh1>|<fsh2>|<fsh3>

DATA LENGTHS:

Part (1)=Fixed at 126 characters

Part (2)=Fixed at 171 characters

WHERE (all parameters eight (8) characters in length):

z r

=

Zero Reference

zl1/2/3

=

Zero low temperature a/d readings.

zh1/2/3

=

Zero high temperature a/d readings.

s r

=

Span Reference

sl1/2/3

=

Span low temperature a/d readings.

sh1/2/3

=

Span high temperature a/d readings

t r

=

Reference temperature

tl1/2/3

=

Low cal temperatures.

th1/2/3

=

High cal temperatures.

fzl1/2/3

=

Zero functions at low temperatures.

fsl1/2/3

=

Span functons at low temperatures.

fzh1/2/3

=

Zero functions at high temperatures.

fsh1/2/3

=

Span functions at hight temperatures.

! TEMPERATURE DIAGNOSTICS DATA RECALL:

COMMANDS: T1R ,T2R and T3R
DATA FORMAT (1):

<zc>|<sc>|<wa>|<ftl>|<ftl2>|<ftl3>|<fth>|<fth2>|<fth3>|<zr>|<zl1>|
<zl2>|<zl3>|<zh1>|<zh2>|<zh3>

DATA FORMAT (2):

<sr>|<sl1>|<sl2>|<sl3>|<sh1>|<sh2>|<sh3>|<tr>|<tl1>|<tl2>|<tl3>|
<th1>|<th2>|<th3>

DATA FORMAT (3):

<fzl1>|<fzl2>|<fzl3>|<fsl1>|<fsl2>|<fsl3>|<fzh1>|<fzh2>|<fzh3>|
<fsh1>|<fsh2>|<fsh3>

LENGTH:

Part (1)= Fixed at 143 characters
Part (2)= Fixed at 125 characters
Part (3)= Fixed at 107 characters

WHERE (all parameters eight (8) characters in length):

z r

=

Zero Reference

zl1/2/3

=

Zero low temperature a/d readings.

zh1/2/3

=

Zero high temperature a/d readings.

s r

=

Span Reference

sl1/2/3

=

Span low temperature a/d readings.

sh1/2/3

=

Span high temperature a/d readings

t r

=

Reference temperature

tl1/2/3

=

Low cal temperatures.

th1/2/3

=

High cal temperatures.

fzl1/2/3

=

Zero functions at low temperatures.

fsl1/2/3

=

Span functons at low temperatures.

fzh1/2/3

=

Zero functions at high temperatures.

fsh1/2/3

=

Span functions at hight temperatures.

16.0 COMMAND CODE ALPHABETICAL (xxD=COMMAND, xxR=RECALL & XxE=ENTRY):

AxE / AxR

= Fill Preact

CPE / CPR

= Capacity.

CxD

= Zero Temp Cold acquire (x=1,2 or 3).

CxD

= Span Temp Cold acquire (x=X,Y or Z).