3 auto set-up menu (absolute), 6 calibration set-up configuration mode selection, Auto set-up menu (absolute) – Condec DLR3110 User Manual

Page 20

Advertising
background image

16

DLR3110 Rack Mount Digital Pressure Calibrator Manual

5.5.3

Auto Set-Up Menu (Absolute)

See Section 9.0 on page 63 for a complete listing of the default settings for all gage and absolute selections under
the AUTO set-up mode.

NOTE: If you continue to press NO key past highest full scale range, DLR3110 will go to exit menu. Hit YES key at exit
menu (see Table 5-3 on page 14).

5.6

Calibration Set-Up Configuration Mode Selection

Use one of the following methods for configuring the calibration set-up mode:

Method 1

CAL (SELECT). See Section 5.4 on page 13 for information on selecting main mode settings.

Method 2

CAL (DIRECT). Closing only rocker switch position

2

on switch

S1

places the DLR3110 directly into

CALIBRATION mode. See Section 5.3 on page 12.

NOTE: If the AUTO SET-UP selection is accepted and completed, the indicator automatically advances to the calibration
mode (see Section 5.5 on page 14).

LED Display

Alpha-Numeric Display

Key Pressed

Remarks

SET

AUTO ?

YES key

See Section 5.5 on page 14

----

GAGE ?

NO key

See Section 5.5 on page 14

—

ABS ?

YES key

See Section 5.5 on page 14

SET-UP

15 PSIA

NO key

Continue to press NO key until the
alpha-numberic display shows the
DLR3110’s full scale range.

YES key

Press YES key when the alpha-numeric
display shows the DLR3110’s full scale
range.

SET-UP

ACCEPT ?

YES or NO

NO key

Continue to press NO key until the
alpha-numberic display shows the
DLR3110’s full scale range. See note
below.

—

—

YES key

Accept data

XXXX

CAPACITY

YES key

“XXX” shown on LED display represents
the full scale range and is based on
previous selection.

—

EXIT ?

YES key

Return to SET-UP/ I/O?. (see Section 5.4
on page 13).

NO key

Return to XXX/CAPACITY

Table 5-5. Auto Set-Up Mode (Absolute) Parameters Selections