13 i/o set-up mode selection – Condec DLR3110 User Manual

Page 27

Advertising
background image

Configuration and Calibration

23

Figure 5-9. Set-Up Mode Selection

5.13 I/O Set-Up Mode Selection

Method 1

I/O (SELECT) - See Section 5.4 on page 13.

Method 2

I/O (DIRECT) - Closing only rocker switch position

3

on switch

S1

only places the indicator directly into the I/O

mode. See Section 5.3 on page 12.

The I/O SET-UP mode accesses the configuration of available inputs and outputs in the DLR. There are five main
groups of configurable inputs and outputs that are defined as follows:

Serial I/O - configures ports 1 and 2 for baud rate, parity, continuous or demand, etc.

Time and date - sets the onboard clock for time and date (allows time and date to be transmitted in the
serial output).

Parallel output - configures DLR3110 for BCD output. This is an option requiring the installation of the
parallel output option board.

Analog output - configures DLR3110 for analog output (4-20 mA or 0-10 VDC). This is an option
requiring the installation of the analog output option board.

Discrete input and outputs - configures two discretes to function as either control output(s) or remote
command input(s).

Press

UP

or

DOWN

arrow keys to step through selections. Press

LEFT

or

RIGHT

arrow keys to change the current

state of the selection. The modes may be exited at any time by pressing the

CE

key. If the

CE

key is pressed, the

prompt

Exit? YES or NO

scrolls continuously until the

YES

or

NO

key is pressed.

I/O ?

ZERO

DISPLAY

TARE

PRINT

YES

NO

CE

ENTER