Connecting to rs232 interface, Connecting to rs485 interface, 3 connection of additional analyzing devices – LumaSense Technologies IS 12-Al/S User Manual

Page 15

Advertising
background image

IS 12 AI & IS 12-AI/S

Controls and Installation 15

Connecting to RS232 interface
The transmission rate (in baud) of the serial
interface is dependent on the length of the cable.
Values between 2400 and 115200 Bd may be set.

The baud rate has to be reduced by 50% when the
transmission distance is doubled (see also 5.10
kBaud (baud rate)).
Typical cable length for RS232 at 19200 Bd is 7 m.

Connecting to RS485 interface
Half-duplex mode:
A1 and A2 as well as B1 and B2 are
bridged in the 12-pin round connector
of the connecting cable, to prevent
reflections due to long stubs. It also
safeguards against the interruption of
the RS485 Bus system should a

connecting plug be pulled out. The
master labels mark the connections on
the RS485 converter. The transmission
rate of the serial interface in Baud (Bd)
is dependent on the length of the
cable. Values between 1200 and

115200 Bd may be set.

The baud rate is reduced by 50% when
the transmission distance is doubled (see
5.10 kBaud (baud rate)). Typical cable

length for 19200 Bd is 2 km.

3.1.3 Connection of additional analyzing devices

Additional analyzing instruments, for example a LED digital display instrument, only needs to be
connected to a power supply and the analog outputs from the pyrometer. Other Instruments,
like a controller or printer, can be connected to the display in series as shown above (total load
of resistance max. 500 Ohm).

yellow

LED digital display

controller

power supply

green

brown

white

230 V ~

24 V DC

°C

writer

K

A

B

C

D

J

G

F

E

H

M

L

1 2 3 4 5

6 7 8 9

DGND (red)

RxD (black)

Pyrometer’s side

cable

PC’s side

Terminator 120 Ohm

A
B
S

K

A

B

C

D

J

G

F

E

H

M

L

K

A

B

C

D

J

G

F

E

H

M

L

K

A

B

C

D

J

G

F

E

H

M

L

D

G

N

D

B

1

A

2

B

2

A

1

D

G

N

D

B

1

A

2

B

2

A

1

D

G

N

D

A

2

B

2

A

1

B

1

Master

Pyrometer 32

e.g. address 31

Pyrometer 2

e.g. address 01

Pyrometer 1

e.g. address 00

TxD (violet)

Advertising
This manual is related to the following products: