9 data format upp (universal pyrometer protocol) – LumaSense Technologies IS 12-Al/S User Manual

Page 33

Advertising
background image

IS 12 AI & IS 12-AI/S

Data format UPP (Universal Pyrometer Protocol) 33

9 Data format UPP (Universal Pyrometer

Protocol)

Via interface and a suitable communication software or via “Test” function of the InfraWin
commands can be exchanged directly with the pyrometer.

The data exchange occurs in ASCII format with the following transmission parameters:
The data format is: 8 data bits, 1 stop bit, even parity (8,1,e).

The device responds to the entry of a command with: output (e.g. the measuring value) + CR
(Carriage Return, ASCII 13), to pure entry commands with “ok” + CR.
Every command starts with the 2-digit device address AA (e.g. “00”). This is followed by 2 small

command letters (e.g. “em” for level of emissivity

e

), finished with CR

This is followed, if necessary for that command, by the ASCII parameter “X”. If this parameter
“X” is omitted, then the device resets with the current parameter.

A „?“ after the small command letters answers with the respective settings (only at setting

commands, not at enquiry commands).

Example: Entry: “00em“ + CR

The emissivity setting (

e

) of the device with the address 00 is returned

Answer: “0970“ + CR means Emissivity = 0.97 or 97.0%

Description

Command

Parameters

Reading temperature

value:

AAms

Output: XXXXX (dec., in

1

/

10

°C or °F)

last digit is the decimal place (88880 = Temp.-Overflow)

Reading temperature

value repeated:

AAmsXXX

XXX = 000…999 (XXX = number of measuring values)

Reading basic

temperature range:

AAmb

Output: XXXXYYYY (hex 8-digit, °C or °F)

XXXX

=

beginning

of

temp.

range

YYYY=

end

of

temp.

range

Reading temperature sub

range:

AAme

same as mb

Reading emissivity:

AAem

Output: XXXX (dez. 0010 … 1000 in ‰)

Emissivity:

AAemXX

XX = (10…99%), 00 = 100% (decimal)

Emissivity:

AAemXXXX

XXXX = (0010 … 1000‰) (decimal)

Exposition time t

90

:

AAezX

X = 0 … 6 (decimal)

0 = intrinsic time constant

1 = 0.01 s

3 = 0.25 s

5 = 3.00 s

2 = 0.05 s

4 = 1.00 s

6 = 10.00 s

External deletion:

AAlx

Simulation of an external deletion contact

Delete

maximum value: AAlzX

X = 0 … 8 (dez.)

0 = Maximum value storage off

1

=

0.01

s

4

=

1.00

s

7

=

external

deletion

2 = 0.05

s

5 =

5.00 s

8 = automatically deletion

3 = 0.25 s

6 = 25.00 s

Setting of temperature

sub range:

AAm1XXXXYYYY

XXXX (hex 4-digit) beginning of temp. range (°C)

YYYY (hex 4-digit) end of temp. range (°C)

Analog output:

AAasX

X = 0 or 1; 0 = 0 to 20 mA

1 = 4 to 20 mA