M345X12 Blackbody Manual
Operation 16
For convenience, the tables below show the expected radiance output vs. temperature (setpoint) for
two popular wavelength regions.
Note: These tables assume the ambient and background are at 23°C and apply ONLY if
your blackbody was calibrated thermometrically.
To force the radiance output of a thermometrically calibrated blackbody to desired temperature,
simply invert the sign of the ERROR DEG.C and add this number to the blackbody controller
setpoint. Again, note that this only applies at ambient and background temperatures of 23C.
3.7 Shut down
Turn off power using the power ON/OFF switch. The controller remembers the last setpoint used.