3 installation, 1 installing the m390s, 2 water and purge gas supply requirements – LumaSense Technologies M390S User Manual

Page 11: 1 water, 2 purge gas

Advertising
background image

M390S Manual

Installation 11

3 Installation

3.1 Installing the M390S

To install the M390S:

1. Allow the M390 to warm-up to room temperature for at least 2 hours after unpacking

and prior to applying power to remove any condensation.

2. Connect the power cable with IEC-309 at the VAC to the power cable supplied with the

M390S.

3. Connect the power cable to a VAC service that matches the power requirement

specification noted in the SPECIFICATION section and on the SPECIFICATION LABEL

located on the rear of the M390S.

4. Connect the inlet water hoses to a 80 psi city or other clean water supply. The Customer

is responsible to provide pressure safety relief valves at inlet and outlet.

5. Connect the outlet water hose to an open drain. Outlet (drain) must not be blocked.
6. Connect the gas purge to a pre-purified Argon gas tank. The Customer is responsible to

provide pressure safety relief valves at inlet.

7. Set the regulator to 15 psi.
8. Turn on the water and check for leaks.
9. Check all connecting liquid, gas and electrical connections for integrity.

Note: The Blackbody Cavity (or tubular heating element) is normally shipped in place.

Note: If the element or quartz tube is damaged or requires replacement, please see
the maintenance section for cavity replacement procedure.

3.2 Water and Purge Gas Supply Requirements

3.2.1 Water

Flow 1 GPM (3.8 l/m), Pressure: 90 PSI (6.3 kg/cm2) maximum
Customer is responsible to provide pressure safety valves at inlet and outlet. Outlet (drain) must

not be blocked.

3.2.2 Purge Gas

High Purity (99.99% pure) Argon. (See Warning Note Below). Customer is responsible to provide
pressure safety valves at inlet.

Warning: DO NOT USE NITROGEN (N

2

) FOR PURGE GAS. AT HIGH TEMPERATURES,

CYNIDE GAS COULD BE GENERATED.

Pressure: Regulated at 15 PSI (1.8 kg/cm2)
Power: 208 or 220 to 240 VAC, (Check Section 2.2 for proper voltage) Single Phase 50/60 HZ,
24 KW 110 Amps, 15kW 70A

Advertising