LumaSense Technologies M390S User Manual

Page 18

Advertising
background image

M390S Manual

Operation 18

Typical Radiance temperature calibration look-up table (LUT) used

To gain access to the custom look-up table:
Refer to the 4 elliptical buttons on the face of the controller and use this procedure.

1. Press and Hold 1 until ACCESS appears. If Access Level and Level 1 appears, skip to

step 3.

2. Press 4 until Level 3 appears.
3. Press 4 until 3 appears. Wait and "pass" appears.
4. Press 1 until Lin16 appears.

5. Press 2. until display shows the Output # that corresponds to the calibration point

shown in the listing on the previous page.

6. Use UP/DWN buttons (3 and 4) to dial in the measured radiance temperature from a

freezing point transfer device or other radiance standard. Adjust Out# only!

7. Press 2 to view/change more correction points.

8. When finished with all calibrations and to lock the controller, press 1 until ACCESS

appears.

9. Press 3 until Level 1 appears.