M300 Manual

Installation and Operation 13

6. Press UP button until J module appears.
7. Press SCROLL button twice until ADDRESS appears.
8. Press UP/DOWN buttons to change address.
9. Press PAGE button until ACCESS appears.

10. Press SCROLL button. Display will show LEVEL 3.
11. Press DOWN button until LEVEL 1 appears.
12. Press the center, round button on top row to go back to the normal display mode.

A CD containing a sample program is included with your order. For more details on
communications, refer to the electronic communication manual also found on the CD.
RS485 Option

RS-485 Two wire is available as an option. The connection scheme is shown below:

RS-485 for multi-drop applications

3.2.5 Shutting Down

Turn off power using the power ON/OFF switch.

Warning: DO NOT UNPLUG THE M300! The blackbody has a temperature

controlled cooling fan that remains active until the enclosure temperature is

below about 140°F. If the power is interrupted by external means, this cooling will
not occur and the devices in the enclosure may overheat. If power fails during high
temperature use, remove the top, rear cover plate to allow better convection cooling
until the blackbody is cool.