2 introduction, 1 description, 2 technical data – LumaSense Technologies M300 User Manual

Page 9

Advertising
background image

M300 Manual

Introduction 9

2 Introduction

2.1 Description

The blackbody calibration source features a unique uniformly heated spherical cavity with a
near ideal emissivity of 0.999 or better. The M300 will deliver any temperature between 200 °C

and 1150 °C. A convenient microprocessor based, self-tuning digital PID controller holds the
selected temperature to within 1 °C, assuring high accuracy calibration of infrared
thermometers, thermal imaging equipment, etc.

2.2 Technical Data

Temperature Range:

200 to 1150 °C or 392 F to 2102 °F

Accuracy:

±0.25% of reading ±1 C when using Calibration Certificate

correction factors

Temperature Resolution:

0.1 C

Stability:

0.5 C per 8 hour period

Source Non-Uniformity:

±0.1 C

Heated Cavity Shape:

Spherical

Aperture Diameter:

2.00 (51 mm)

Emissivity :

0.9999 ± 0.0005 effective at 1.5µm to 0.65 µm (radiance
calibration)

Method of Control:

Digital self-tuning PID controller

Warm-up Time:

1 hour from ambient to 1000 C

Operating Ambient

Temperature:

0 C to 44 C or 32 F to 110 F

Cooling:

Fan cooled, air inlet on rear panel

Power Requirements:

208/220 VAC±10% 50/60Hz 11A Maximum

Weight:

175 lbs (80 kgs)

Advertising