Assuming that the monitor and the PC are connected and the
software is running (displaying the Main Menu), the zero calibration
continues as follows:
1. Ensure that the monitor has been switched on for at least 30
minutes.
2. When the Main Menu screen is displayed, select Zero / O
2
Span
from the Bench Command pull-down menu.
When the Multi Gas Monitor – INNOVA 1316A-2 receives the Zero
command, it checks the system status.
When all conditions for zero calibration are met, the Process In
).