8. Servicing/recalibration

Have the scale recalibrated by authorised persons in accordance with national legal re-
quirements. The year of initial calibration can be found behind the CE symbol on the rat
ing plate above the number of the appointed office, 0109 (Hessen Weights & Measures
Office). Recalibration is essential if one or more security marks are damaged.
We recommend having your scale serviced by your local service agent before having it
recalibrated. seca Customer Service will be pleased to give you further assistance.

Changing the batteries

When

or

bAtt

appear in the display, you should change the batteries. You need

6 type AA 1.5 V batteries.
– Carefully turn over the scale so that the

base of the scale is accessible.

– Press the closure of the battery compart-

ment in the direction of the cover itself and
open the battery compartment.

– Remove the used batteries and insert new

batteries.

Check that the polarity is correct.

– Close the cover.

– Then turn the scale back up the right way.

9. Technical data

Maximum capacity

200 kg

Minimum load

2 kg

Graduation
0 to 150 kg

100 g

150 kg to 200 kg

200 g

Accuracy
0 to 50 kg

±50 g

50 kg to 150 kg

±100 g

150 kg to 200 kg

±200 g

Height of figures

20 mm

EN 60 601-1:

medical electrical equipment type B

Dimensions
Width:

321 mm

Height:

60 mm

Depth:

360 mm