Setting up and aligning the scale – Seca 899 User Manual

Page 22

Advertising
background image

22

Setting up and aligning the scale

The screwed feet can be used to level off

any slight unevenness in the floor.
– Place the scale on a level surface.
– Align the scale by screwing the foot

screws. The air bubble in the spirit level

must be right in the centre of the circle.

– Ensure that only the feet of the scale are

in contact with the floor. The scale may

not be in contact at any other point.

Important!

The alignment of the base of the scale

must be checked and corrected if nec-
essary every time the location of the

scale is changed.

spirit level