Model 869

25

GB

6. What do I do if…

… no weight display comes on?
– Is the scale switched on?
– Check the batteries.
– Is the plug contact for connecting the display to the scale correctly connected?
– Is the plug contact of the mains unit properly connected to the scale?

0.0

does not appear before weighing?

– Switch off the scale with the Start key and start the scale again – there must not be

any load on the scale – and only its feet should be in contact with the floor.

… one segment is illuminated constantly or not at all?
– The relevant segment has a fault. Inform Servicing.

… the display shows

?

– Battery voltage is running low. Change the batteries in the next few days.

bAtt

appears in the display?

– Batteries are empty. Put in new batteries.

St0p

appears in the display?

– Maximum load has been exceeded.

… the display flashes?
– If you have just activated the Hold or Tare function, please wait until the scale has de-

tected a steady weight value. The display will then stop flashing.

– If there has been no activation of a function beforehand, remove the load from the scale

and wait until

0.0

is displayed, then weigh again.

… the display

E

and XX.XX appear?

– Switch off the scale with the Start key and start the scale again. The scale will then

work normally again.

If this is not the case, disconnect the power supply by briefly removing the batteries. If

this measure is equally unsuccessful, inform the Service department.

7. Servicing

Your seca scale leaves the factory with an accuracy greater than ± 0.15 %. To ensure that

this accuracy is maintained, the product must be carefully set up and regularly maintained.

Depending on the frequency of use, we recommend servicing every 3 to 5 years. Please

ensure that a qualified service agent is used. If in doubt, contact seca Customer Service for

further information.