Operation, 1 making the scale operational, Moving the scale – Seca 677 User Manual

Page 69: Aligning the scale

Advertising
background image

Before you really get started ... • 69

English

6. OPERATION

CAUTION!

Bodily injury

Perform a function check as described in the
section entitled "Function check" on page 87
before using the device each time.

6.1

Making the scale operational

Moving the scale

1. Disconnect the power pack connector from the

power supply socket.

2. Disconnect the power pack connector from the

connecting socket on the scale.

3. Undo the locking screws and carefully fold down

the railing until it abuts completely.

4. Fold up the latch until it engages securely in the

railing.

5. Fold away the access ramps.

6. Carefully align the device with the railing.

7. Move the device in this position to the desired

setup or storage location.

Aligning the scale

ATTENTION!

Incorrect measurement due to force shunts

If the scale and housing is in contact with some-
thing, e.g. a towel, weight will not be measured
correctly.
– Set the scale up so that only its foot screws

are in contact with the floor.

1. Place the scale on firm, level surface.

2. Undo the knurled wheels.

3. Level the device by turning the foot screws.

The air bubble in the spirit level must be located in
the exact center of the circle.

Latch

Locking screws

2.

3.

4.

Advertising
This manual is related to the following products: