IV. CALIBRATION

Calibration is NOT required before first use. Only

calibrate if scale is not weighing properly.

**Make sure you use the correct calibration weight for

your scale or the calibration will not be accurate**

AMW-150DR: 50g + 100g calibration weight

AMW-300DR: 100g + 300g calibration weight

A. AMW-150DR Calibration

1. Press the [ON/OFF] key to turn on the scale.

2. Once the scale goes to zero (0.0 or 0.00g) press the

[CAL] key once. The LCD will display “50.00g”.

3. Press the [MODE] key once. The display will now

being to flash “50.00g”.

4. Place the 50g weight on the platform, and press

the [MODE] key again. The display will now show

“100.00g”. Tap [MODE] again to make the display flash

“100.00g”.

6. Place another 50g on the platform to make 100g

total on the scale and press the [MODE] key. The

display will show PASS and shut off. Calibration is

compete!

B. AMW-300DR Calibration

1. Press the [ON/OFF] key to turn on the scale.

2. Once the scale goes to zero (0.0 or 0.00g) press the

[CAL] key once. The LCD will display “50.00g”.

3. Press the [MODE] key once. The display will now

being to flash “50.00g”.

Advertising