2. Technical Data

2.1

Humidity Detection

Ranges are possible from 0-100 %RH
as well as absolute Humidities, Dewpoints
and Enthalpie, programed is a range of

...........................

Resolution

0,1 % RH

Sensing Element (capacity type)

SE – 02

Influence of Temperature on the Humidity

± 0,01 % RH / °C

Repeatability

< 0,5 % RH

Hysteresis for 4 hours Cycle
10 % - 95 % - 10 % RH

< 0,5 % RH

Unlinearity of Electronics / Humidity

< 1 % RH

2.2

Temperature Detection

Ranges are possible from –50°C to

+150°C, programed range is

..................°C

Resolution

0,1 °C

Temperature sensing Element

Pt - 1000

System Accuracy of Temperature

± 0,35 °C

Unlinearity of Electronics / Temperatur

< 0,15 °C

2.3

Power Supply

Standard Power Supply

90-240 VAC,
47-63 Hz

or Low Voltage Supply

10-30 VDC/9-25 VAC

2.4

Maximum Operational Temperatures

Transmitter Electronics

-20....+40 °C

Probe Electronics

-40....+60 °C

Probe Tube with Sensors

-40....+120 °C

2.5

Power Consumption

Max. Power Consumption

approx. 5 VA

HYGROCONTROL Manual for Type 86 Ex Page

5