COSA Xentaur Xentaur LPDT User Manual

Page 13

Advertising
background image

LPDT User’s Manual

page 6

Under a pressurized system removal of the sensor with-
out the installation of isolation valves can be dangerous.

If in-situ installation is required, bypass mounting is prefer-
able; make sure to install the sensor at the upper surface of
the gas line to minimize its exposure to liquid water, should
condensation occur, the XTR65W sensor is best suited for
these applications. Also consider the need to isolate (depres-
surize) before installing or removing the sensor.

Main

Gas Line

LPDT

LPDT

4-20mA loop cable

Safety shut-off Valve

Bypass
Control

Valve *

Safety shut-off Valve

Bypass Installation, Sensor

Measuring at Line Pressure

In-Line Installation, Sensor

Measuring at Line Pressure

Main

Gas Line

4-20mA loop cable

oint Transmitter

Ma

Se

Mo

Think
Safety

Read

Man

ual

www.xentaur.com

NOT RECOMMENDED

Sample Cell

oint Transmitter

Ma

Se

Mo

Think
Safety

Read

Man

ual

www.xentaur.com

* maintain differential pressure to provide adequate flow through sample cell

Advertising