COSA Xentaur Xentaur LPDT User Manual

Page 36

Advertising
background image

Section 3: Instrument

page 29

Legend:

denotes alternately flashing messages.

DISPLAY

EXPLANATION

REQUIRED ACTION

LO

(55

PROM check sum failed.

pow

er

on t

es

ts

HI

(55

RAM write/read test failed.

cycle power

%

(55

Unidentified power-up failure.

if problem persists,

&6)

EEPROM Check Sum Failed.

return to your representative

(55$'

A/D converter failure.

s

y

st

em

t

es

ts

onc

e pe

r 2 m

in.

for service.

(555()

Reference voltage for A/D out of spec.

/2%$7

Low supply voltage.

check input voltage 10-33VDC

LO

51*

TEMP

Instrument low temperature range

has been exceeded.

make sure that the unit is at a

HI

51*

TEMP

Instrument high temperature range

has been exceeded.

temperature of -10°C to +70°C.

231

Sensor circuit is open.

de

w

poi

nt

di

sp

la

y t

e

st

s

see table in

6$7

Sensor is saturated.

sensor troubleshooting

6+5

Sensor circuit is shorted.

section



Trying to calculate dewpoint for

undefined sensor.

select sensor and

SpanCheck.

'%

Trying to calibrate an undefined

sensor.

Ca

li

bra

ti

on. t

es

ts

see Calibration instructions.

6(1/2

Sensor reading is ‘too’ low to be a satu-

rated sensor, for SpanCheck.

(55((3

EEPROM write cycle not completed.

m

is

c

el

la

ne

ous

if this persists, return for service.

/2&

Attempting to modify a locked unit.

unlock unit, see set-up mode.

51*

number can not be displayed in 3.5 digits.

°C(°F)

w/flashing

DEWPOINT

An XTR-65 sensor is measuring less

than -70°C or -94°F dewpoint.

PSI

flashing

when

dew

Pressure correction is enabled.

;(1

Turn on message, Xentaur (Greek

Ξ

=X)