2 electrical connections, 1 connecting power, 2 electrical connections 3.3.2.1 connecting power – COSA Xentaur Xentaur XDT User Manual

Page 19

Advertising
background image

XDT User’s Manual

12

3.3.2 Electrical Connections

All connections are made via a 21 contact, pluggable screw terminal block referred to as P1.

Cable access to the XDT-NEMA enclosure is through three ports (refer to appendix E), which
maintain the NEMA integrity:

One 1/2” NPT conduit fitting, may be used for high or low voltage connections. This port is
intended for the power and high voltage relay wiring.

Two watertight cable grips located at the bottom of the instrument: for low voltage wiring
only
such as sensor, analog output, RS-232, or relays when used with low voltages. They
accommodate cables with diameters 0.196” to 0.315”. To install a cable, loosen the nut, feed
the cable through the grip and tighten the nut again. Unused cable grips should be plugged to
maintain the integrity of the enclosure.

3.3.2.1 Connecting Power

The XDT is typically provided to be field wired to operate from AC Mains Power of 100 to
250VAC, 50/60Hz, 10VA, the instrument has an internal 0.5A, 250V fast acting fuse. However
the XDT can also be optionally ordered to operate from 15 to 30 VDC 0.5A, with an internal 0.5A

21 C

20 NO

18 Not Connected

14 Not Connected

3

Not Connected

19 NC

17 C

16 NO

15 NC

ALARMS

*

LO

HI

Ordinary relay contacts: 10A, 250VAC or 30VDC
Hermetically Sealed relay contacts: 3A, 120VAC

11 RET

10 RxD

9

TxD

8

A-RET

7

A-OUT

2

AC NEUTRAL

1

AC LIVE

4

AC GROUND

13 AC GROUND

AC Power Input:
100 to 250VAC, 50/60Hz, 10VA

Analog Output 4/20mA or 0/24mA or Voltage

*

SENSOR CABLE

+DC power 15V OUTPUT and/or 15-30V INPUT

*

0.5A

5

SIG-IN

6

SIG-RET

12 PS+

white

black

0.5A 250V

0.1uf

0.1uf

0.0047uf

0.0047uf

4.3mH

RFI filter board

275VAC

Power Supply Module

Sensor
Measuring
Circuitry

1M Ohm

33V 400W

Logic ground and AC Power

ground are connected for safety

and electromagnetic

interference considerations

Instrument Circuit board

Pluggable Block Terminals

RS 232 I/O

*

Note:

*

Options will function only if they are installed

CAUTION: Connecting the RS-232C interface
on instruments which do not have the option
installed, will damage the instrument.

Advertising