Xdt user’s manual appendices 36 – COSA Xentaur Xentaur XDT User Manual

Page 43

Advertising
background image

XDT User’s Manual Appendices

36

Flow Diagram Operating State User Interface (continued) Manual & Scheduled Self Calibration

˚C

DEWPOINT

HI/LO

SET

SET

SET

SET

SET

SET

PSI

NUMBERS

CALCULATING

CALIBRATION

Legend: A slash '

/

' in the display area, is used to depict two alternately shown (flashing back and forth) messages.

denotes beeping. Buttons without designator or arrow leading out, perform no function

LOC

-65.0

CAL

XSN/123

END

SEN/2LO

AC/_AC

CNF/AC?

CAN

30 seconds anywhere (other than
"Calibration") without key press

appears when changing
settings of locked unit

Viewing Dewpoint

Start

Calibration

Viewing

Serial-

number

Power ON initialization
and sign on message

Flashing PSI indicates
pressure corrected
dewpoint

Flashing HI and/or LO

indicates corresponding

relays not energized

SET

SET

SET

SET

˚C

DEWPOINT

˚C

DEWPOINT

SLF/CAL

-42.0

720/HRS/INTR

150/HRS/LEFT

SET

DIS

BEG/S.C.

CNF/SC?

SLF/CAL/-41.5

˚C

DEWPOINT

S.C,/STB/-41.5

˚C

DEWPOINT

S.C./FIN/-42.0

{

values +/-0.1

˚

C

{

values +/-1hr.

{

timeout

timeout

cancel

cancel

abort performing Self Cal Now

disable Self Cal corrections

Scheduled Time Elapsed,
Entry Point

dew is stable

perform
Self Cal

During Self Cal the Analog output and Alarm Relays are frozen to their previous states

Self Cal

Completed go to

display Dewpoint

Confirm Entry into
Self Cal Menus

Set Scheduled Calibration
Interval in hours

0 (blank)=Scheduled SelfCal disabled,

but SelfCal corrections not affected

View Hours left until
next Scheduled
Calibration

The Self Cal Corrections
will not be used in dew
computations from now on,
until another Self Cal is
performed, manually or
scheduled. Schedules
are not affected.

Set Reference
Gas dewpoint

Begin Self Cal Now

Electric Valve Activation, and
15 min. time out for system stabilization,
performed only if electric valve is installed.
Displays dewpoint being measured now.

Dewpoint stable testing,

0.5

˚

C for 3 min. Displays

dewpoint being measured now.

Electric Valve Deactivation, and final
Dewpoint stable testing, 0.5

˚

C for 3 min.

performed only if electric valve is installed.
Displays dewpoint being measured now.

Confirm performing
Self Cal NOW

Appears only if

unit is equipped

with electric valve

{

units

Toggle Pressure Correction

SHORT PRESS

1/2 sec

{

{

Sensor

Reading

New calibration
values are saved

Cancel Calibration (it is not performed)

Confirm SpanCheck

Confirm
SpanCheck
again

1 minute

few sec

1 sec

OK

TOO LOW

Press together
to RESET the
instrument

go to
Set-Up
State

ignore error

When ALARM OPTIONS are installed men

us appear here

Flow diagram items not related to Self Calibration are shaded in gray.

Advertising