Xdt user’s manual 30, Display explanation required action, Display is dark – COSA Xentaur Xentaur XDT User Manual

Page 37

Advertising
background image

XDT User’s Manual

30

Legend: 

denotes a beeping accompaniment to the message.

/

denotes alternately flashing messages.

DISPLAY

EXPLANATION

REQUIRED ACTION

LO ERR



PROM check sum failed.

power

on

tes

ts

HI ERR



RAM write/read test failed.

cycle power

% ERR



Unidentified power-up failure.

if problem persists,

C.S.F/



EEPROM Check Sum Failed.

return to your representative

ERR/AD



A/D converter failure.

s

y

st

em

tes

ts

once per

2

m

in.

for service.

ERR/REF



Reference voltage for A/D out of spec.

LO/BAT



Low supply voltage.

LO RNG/



TEMP

Instrument low temperature range

has been exceeded.

make sure that the unit is at a

HI RNG/



TEMP

Instrument high temperature range

has been exceeded.

temperature of -10°C to +50°C.

OPN/



Sensor circuit is open.

dewpoint

di

spl

ay

te

sts

SAT/



Sensor is saturated.

see table in sensor troubleshooting section 2.5

SHR/



Sensor circuit is shorted.

...

Trying to calculate dewpoint for undefined sensor.

select sensor and recalibrate.

DB



Trying to calibrate an undefined sensor.

ca

libra

ti

on

te

sts

see calibration instructions.

SEN/2LO



Sensor reading is ‘too’ low to be from a saturated sensor,

thus can not be calibrated

2LO

While Compensating for cable length, the measurement did not

correspond to the expected open or short values.

cabl

e

te

sts

Check the means of opening & shorting the sensor

cable. see section 3.4.5 - 11

2HI

ERR/EEP



EEPROM write cycle not completed.

mis

ce

lla

ne

ous

if this persists, return for service.

LOC



Attempting to modify a locked unit.

unlock unit, see set-up state 3.4.5.-10.

SEC



Attempting to unlock a secured unit.

HI (flashing while
viewing DEWPOINT)

The HI Alarm (alarm #1) relay is deenergized.

See section 3.4.4.2 Setting the Alarms.

LO (flashing while
viewing DEWPOINT)

The LO Alarm (alarm #2) relay is deenergized.

A.O./OPN



The Analog Output ‘sees’ an open circuit

check the cables connecting the analog output, make

sure the terminating impedance is less that 500

Ω.

PSI flashing when dis-
playing dewpoint

Pressure correction is enabled, the dewpoint is computed using

programmed pressures

refer to section 3.4.4.1

-70°C (-94°F) with
flashing DEWPOINT

An XTR-65 sensor is measuring less than -70°C or -94°F D.P.

check instrument

upgrade to XTR-100 sensor

units flash and revert
back to original state

The user changed the units of measure when the analyzer is locked

the software keeps analog output with original inits

unlock analyzer, change units, lock analyzer

(see section 3.4.4.1)

RNG

number can not be displayed in 3½ digits

XEN



Turn on message, Xentaur (Greek

Ξ=X)

display is dark

No power to instrument

check power to instrument, check fuse