Connecting power, Interfacing to analog output – COSA Xentaur Xentaur XDTL-HT User Manual

Page 13

Advertising
background image

Installation

13

Connecting Power

The XTDL-HT™ requires AC power of 110/230VAC, 50/60 Hz. The voltage rating should be specified on
the order and it is marked on the front panel label. Each instrument is provided with an international
approved AC power cable, terminated with a connector according to the ordered option. This cable is
provided strictly for testing the instrument; it should not be used in the final installation. For
installation, the instrument must be field wired permanently connected and installed according to local
and nationally recognized procedure for equipment of this type and stated power requirements. For
North American installations, use UL 62 and/or CSA C22.2 No 49, 16AWG cable with black, white, and
green color codes. For European and other international installations, use CENELEC harmonized type
cable, with 1.31 mm² wire size or equivalent, with brown, light blue, and green/yellow stripe color
codes. Power cable must be wired to the DIN-mounted terminal blocks marked 1 (AC live In), 3 (AC
neutral In), and 5 (AC ground). The cable must be fed through one of the cable grips (second cable grip
is recommended).

Interfacing to Analog Output

The optional analog outputs can be voltage or
current depending on the ordered configuration. The
analog output cable should be fed through one of
the cable grips (fourth cable grip is recommended).
Channel 1 output should be wired to the DIN-
mounted terminal blocks marked 14 (positive) and
15 (negative). Isolation option on the analog outputs
is available upon request. Refer to

Analog Output

chapter for setup information.

CAUTION:

Do not connect to AC power without a proper grounding connection.

Analog output cable installation

North American power cable installation

European and other international power cable installation