Appendix B: Modbus Registers

Register
Number

Address

Access

Level

Data
Type

Description

Values

40668

667 (29Bh)

User

16-bit

TCP port

Valid values are 0 - 65535

40669

668 (29Ch)

Admin

48-bit

EUI-48 MAC Address

Beeper Registers

40791

790 (316h)

User

16-bit

Master volume

Valid values are 0 - 100%

Display Registers

40801

800 (320h)

User

16-bit

Screen saver time

0=Off
1=5 minutes
2=10 minutes
3=30 minutes
4=1 Hour
5=2 Hours
6=4 Hours

40802

801 (321h)

User

16-bit

Display enable time

Valid values are 0 - 2359

40803

802 (322h)

User

16-bit

Display disable time

Valid values are 0 - 2359

40804

803 (323h)

User

16-bit

Contrast

Valid values are 0 - 100%

Reading Registers

40831

820 (334h)

User

16-bit

Primary reading variable

0=Dewpoint
1=Dewpoint in °C
2=Dewpoint in °F
3=Dewpoint in ppmV
4=Dewpoint in lbs H

2

O/mm scf

5=Dewpoint in g/m

3

6=Sample temperature
7=Sample pressure
8=Enclosure pressure
9=User input 1
10=User input 2
11=User input 3
12=User input 4

40832

821 (335h)

User

16-bit

Secondary reading variable

See primary reading variable