Troubleshooting The I/R Link:

Most problems tend to occur due to the positioning of the meter relative to the

Base Unit. There are a few error messages that may appear on the meter’s

screen:

a. Timeout: This error message occurs when too much time has passed

between data sent by the meter and the response that should have been

received from the Base Unit. Possible causes:

o

Meter misalignment with the Base Unit (make sure the

communication windows are facing each other, within the distance

indicated).

o

No connection to the PC. The Base Unit must be connected to the

PC through a USB cable. Check connection.

b. KBD: This message occurs when the user presses a key during

communications. The meter interprets this as a request to escape this activity

and returns to the Main Menu. The Base Unit should return to a green LED

status in a few seconds. A message should appear on the PC screen as well,

indicating lost communications.

c. Sync: This message occurs when data framing is lost during data

transmission. This would only happen if the wireless link becomes unreliable.

Ensure the meter and Base Unit are aligned and restart the session.

d. Data E: This message occurs when the content of a correctly framed data

packet is invalid. Again, this is an unreliable wireless link problem.

NOTE: Individual records cannot be downloaded to the PC. The entire storage

content is delivered to the PC during a link session.

SETUP MENU:

The Setup group of features allows access to the configurable parameters of the

meter. They are: Date and Time, Power-off time, Backlight, and Conversions