Configuring the edge trigger – GW Instek GDS-1000-U Series User Manual User Manual
Page 87

CONFIGURATION
87
>
Longer than =
Equal to
<
Shorter than ≠
Not equal to
Trigger slope
Triggers on the rising edge.
Triggers on the falling edge.
Trigger coupling AC
Triggers only on AC component.
DC
Triggers on AC+DC component.
Frequency
rejection
LF
Puts a high-pass filter and rejects the
frequency below 50kHz.
HF
Puts a low-pass filter and rejects the
frequency above 50kHz.
Noise rejection
Rejects noise signals.
Trigger level
LEVEL
Using the trigger level knob moves the
trigger point up or down.
Configuring the edge trigger
Procedure
1. Press the Trigger menu key.
MENU
2. Press Type repeatedly to
select edge trigger.
Type
Edge
3. Press Source repeatedly to
select the trigger source.
Source
CH1
Range
Channel 1, 2, Line, Ext
4. Press Mode repeatedly to
select the Auto or Normal
trigger mode. To select the
single trigger mode, press
the Single key.
Mode
Auto
SINGLE