Probe compensation, Gds-1000a-u series user manual, Press the utility key – GW Instek GDS-1000A-U Series User Manual User Manual
Page 132: Press probecomp, Menu

GDS-1000A-U Series User Manual
132
7. The Channel1 calibration
will complete in less than 5
minutes.
Ch1 calibration 1/3
8. When finished, connect the
calibration signal to the
Channel 2 input and repeat
the procedure.
Done!!
9. When the calibration is complete the display
will go back to the previous state.
Probe Compensation
Procedure
1. Connect the probe between the Channel1 input
and the probe compensation output (2Vp-p,
1kHz square wave) on the front panel. Set the
probe voltage attenuation to x10.
VOLTS/DIV
VOLTS/DIV
TIME/DIV
CH 1
MATH
CH 2
MENU
MENU
Acquire
Display
Utility
Help
Run/Stop
VARIABLE
FORCE
Autoset
Cursor
SINGLE
Hardcopy
Measure
Save/Recall
LEVEL
VERTICAL
HORIZONTAL
TRIGGER
CH1
CAT
300V
M
W
15pF
MAX. 300Vpk
1
CH2
EXT TRIG
CAT
300V
M
W
15pF
MAX. 300Vpk
1
X
Y
X
1
0
X
1
CH1
x1
x10
2. Press the Utility key.
Utility
3. Press ProbeComp.
ProbeComp
Menu