Load wire assembly – GW Instek PSH-A Series 360W User Manual User Manual
Page 15

Setup
29
Series Operation (cont.)
Series
Operation +
Remote
Sensing
PSH #1
+OUT
+IN
Load
+Sense
PSH #2
+OUT
+Sense
—IN
—OUT
—OUT
—Sense
—Sense
First PSH:
Negative output → Negative load input
Positive output → The next PSH negative output
Negative sense → Negative load input
Positive sense → Positive output monitor
Intermediate PSH:
Negative output → The previous PSH positive
output
Positive output → The next PSH negative output
Negative sense → The previous PSH positive
output
Positive sense → Positive output monitor
Last PSH:
Negative output → The previous PSH positive
output
Positive output → Positive load input
Negative sense → The previous PSH positive
output
Positive sense → Positive load input
PSH
User
Manual
30
Load Wire Assembly
1. Wire
selection
Select the appropriate wire according to the
guideline on page25.
2. Terminal
screw
replacement
Replace the two screws on the output terminal
with the hex screw in the output connection kit.
3. Terminal
connection
Screw load wires to the terminal.