Ovp calibration – GW Instek PSH-A Series 360W User Manual User Manual
Page 34

Calibration
67
5. Adjust the display value to the
Multimeter / Current Shunt
reading.
Wheel knob
changes the value.
F/C key
F/C
moves the cursor
before and after the decimal point.
Multimeter reading: 0.02A
Calibration
CurrMin 0.02A
6. Press the Vset/Iset/Enter
key
V
SET
/ I
SET
ENTER
to confirm the setting.
The display goes back to the
default calibration menu.
Calibration
Current
7. Save the result and exit the
calibration mode (Save), OR
Exit the calibration mode without
saving the result (Exit).
Calibration
Save
Calibration
Exit
PSH
User
Manual
68
OVP Calibration
Connection
No connection required
Calibration step
1. Enter the Voltage calibration
menu. See page63 for details.
Calibration
OVP
2. Press the Vset/Iset/Enter
key
V
SET
/ I
SET
ENTER
. PSH calibrates OVP
automatically.
Calibration
OVP
100%
3. The display goes back to the
default calibration menu.
Calibration
OVP
4. Save the result and exit the
calibration mode (Save), OR
Exit the calibration mode without
saving the result (Exit).
Calibration
Save
Calibration
Exit