GPD-X303S Series User Manual
Load Cable Connection
GTL-104A
1. Turn the terminal
counterclockwise and loosen
the screw.
2. Insert the cable terminal.
3. Turn the terminal clockwise
and tighten the screw.
GTL-105A
Insert the plug into the socket.
GTL-203A, 204A
Insert the plug into the terminal.
Wire type
When using load cables other than the attached,
make sure they have enough current capacity for
minimizing cable loss and load line impedance.
Voltage drop across a wire should not excess 0.5V.
The following list is the wire current rating at
450A/cm
2
.
Wire size (AWG)
Maximum current (A)
20
2.5
18
4
16
6
14
10
12
16
24