Rcl<nr1, Sav<nr1, Baud<nr1 – GW Instek GPD-4303S User Manual
Page 51: Local
Advertising

REMOTE
CONTROL
Return parameter
GW INSTEK,GPD-X3303,SN: xxxxxxxx, Vx.xx
(Manufacturer, model name, serial number,
firmware version)
RCL<NR1>
Description
Recalls a panel setting.
<NR1>
1 – 4: Memory 1 to 4
Response time
Minimum 10ms
Example RCL1
Recalls the panel setting stored
in memory 1
SAV<NR1>
Description
Stores the panel setting.
<NR1>
1 – 4: Memory 1 to 4
Response time
Minimum 10ms
Example SAV1
Stores the panel setting in
memory 1
BAUD<NR1>
Description
Sets the baud rate to 9600bps/57600bps
/115200bps.
<NR1>
0: 115200bps, 1: 57600bps, 2: 9600bps
Response time
Minimum 10ms
Example BAUD0
Sets the baud rate to 115200bps.
LOCAL
Description
Exits remote mode and sets the instrument to local
mode.
51