Technical specifications, Sirius89n, En - 100 – HT instruments SIRIUS89N User Manual
Page 102

SIRIUS89N
EN - 100
14. TECHNICAL
SPECIFICATIONS
14.1. TECHNICAL
FEATURES
Accuracy is indicated as [% of reading + number of digits]. It refers to the following
atmospheric conditions: a temperature of 23°C ± 5°C with a relative humidity < 60%.
14.1.1. Safety Test functions
LOW
Ω: 200mA CONTINUITY TEST (AUTO, RT+, RT- MODE)
Range [
Ω] Resolution
[
Ω]
Accuracy(*)
0.01
ч 9.99
10.0
ч 99.9
0.01
0.1
±(2% Reading + 2 digit)
(*) After Test leads calibration
Test Current
> 200mA DC per R
≤5Ω (Test leads included)
Resolution for Test current:
1mA
Open Circuit Voltage
4V
≤ V0 ≤ 24V
M
Ω: INSULATION TEST
Test Voltage [V]
Range [M
Ω] Resolution
[M
Ω]
Accuracy
0.01
ч 9.99
0.01
10.0
ч 49.9
0.1
±(2% Reading + 2 digit) se V/R>1µA
50
50.0
÷ 99.9
0.1
±(5% Reading + 2 digit) se V/R≤1µA
0.01
ч 9.99
0.01
10.0
ч 99.9
0.1
±(2% Reading + 2 digit) se V/R>1µA
100
100.0
÷ 199.9
0.1
±(5% Reading + 2 digit) se V/R≤1µA
0.01
ч 9.99
0.01
10.0
ч 199.9
0.1
200
ч 249
1
±(2% Reading + 2 digit) se V/R>1µA
250
250
÷ 499
1
±(5% Reading + 2 digit) se V/R≤1µA
0.01
ч 9.99
0.01
10.0
ч 199.9
0.1
200
ч 499
1
±(2% Reading + 2 digit) se V/R>1µA
500
500
÷ 999
1
±(5% Reading + 2 digit) se V/R≤1µA
0.01
ч 9.99
0.01
10.0
ч 199.9
0.1
200
ч 999
1
±(2% Reading + 2 digit) se V/R>1µA
1000
1000
÷ 1999
1
±(5% Reading + 2 digit) se V/R≤1µA
Open circuit Test Voltage
<1.3 x Nominal Test Voltage
Short Circuit Current
<6.0mA with 500V Test Voltage
Nominal Test Current
500V
>2.2mA with 230k
Ω
other
>1mA with 1k
Ω*Vnom
RCD: TEST ON RCD DEVICES
Nominal Test Current (I
∆N)
10mA, 30mA, 100mA, 300mA, 500mA
RCD type
AC, A General e Selective
Phase to Earth Test Voltage
100V
÷ 250V
Frequency 50Hz
± 0.5Hz
- Tripping Time Measurement t
∆N
Range [ms]
Resolution [ms]
Accuracy
½ I
∆N
, I
∆N
1
÷999
2 I
∆N
1
÷200
general
1
÷250
selective
5 I
∆N
RCD 1
÷50
general
1
÷160
selective
1
±(2%Reading+2digit)
- Contact Voltage Ut
Range [V]
Resolution [V]
Accuracy
0
÷ 2U
t lim
0.1
- 0%, +(10% Reading + 3digit)
Ut LIM (UL): 25V o 50V
- Global Earth Resistance Measurement (avoiding RCD tripping)
Range [
Ω] Resolution
[
Ω] Accuracy
I
∆N
1
÷ 1999
1
±(5% Reading + 3digit)
Test Current
0.5 I
∆N set
- Tripping Current Measurement
I
∆N
RCD Type
Range I
∆N
[mA]
Resolution [mA]
Accuracy I
∆N
AC
(0.5
÷ 1.4) I
∆N
I
∆N ≤ 10mA
A
(0.5
÷ 2.4) I
∆N
AC
(0.5
÷ 1.4) I
∆N
I
∆N > 10mA
A
(0.5
÷ 2) I
∆N
0.1 I
∆N
- 0%, +10% I
∆N
FREQUENCY
MEASUREMENT
Range [Hz]
Resolution [Hz]
Accuracy
47.0
÷ 63.6
0.1
±(0.1%Reading+1 digit)
RCD and LOOP function are active only for 50Hz
± 0,5Hz frequency