HT instruments SIRIUS89N User Manual
Page 95

SIRIUS89N
EN - 93
START-UP
ANALYZER CONFIG:
All the parameters:
unmodified
RECORDER CONFIG:
Start:
MANU (the recording starts at
00 sec mark on clock after
pressing the START/STOP key)
Stop: MANU
Integration period:
5sec
Recording of harmonics:
ON
Recording of Voltage anomalies (voltage Dips and Surge) :
ON
Voltage Reference for Dips and Surge detection:
230V
Upper Limit for Dips and Surge detection:
6%
Lower Limit for Dips and Surge detection:
10%
Selected voltages:
V1
Selected voltage harmonics:
THD, 01, 03, 05, 07, 09, 11, 13, 15
Selected currents:
I1
Selected current harmonics:
THD, 01, 03, 05, 07, 09, 11, 13, 15
CO-GENERATION: OFF
Selected powers, Pf and cos
ϕ:
P1, Q1i, Q1c, S1, Pf1, dpf1
Energies:
Ea1, Eri1, Erc1
POWER & ENERGY
ANALYZER CONFIG:
All the parameters:
unmodified
RECORDER CONFIG:
Start:
MANU (the recording starts at
00 sec mark on clock after
pressing the START/STOP key)
Stop: MANU
Integration period:
15min
Recording of harmonics:
OFF
Recording of Voltage anomalies (voltage Dips and Surge) :
OFF
Selected voltages:
V1
Selected currents:
I1
CO-GENERATION: ON
Selected powers, Pf and cos
ϕ:
P1, Q1i, Q1c, S1, Pf1, dpf1
Energies:
Ea1, Eri1, Erc1
By pressing
START/STOP the recording of the selected parameters is started according to
the settings made for each Typical Configuration. The rotary switch position doesn’t affect
the recording setting.