In-Situ pH/ORP Sensor User Manual

Page 4

Advertising
background image

Page 4

• Readings during calibration at pH 7 are greater

than +30 mV or less than -30 mV.

• Sensor is slow to respond.
Note: If the sensor fails to calibrate after you replace the fill

solution, replace the reference junction.

Replacing the Filling Solution
1. TROLL 9500—Use the sensor removal tool to pry

the sensor up and out of the instrument. Aqua
TROLL 400—Grasp the sensor by the finger grip
and pull out of the instrument.

2. Unscrew the reference junction. See Figure 2.

3. Protect the connector end of the sensor with

the dust cap it shipped with, or wrap the sensor
in a paper towel to prevent filling solution from
entering the electrical connector.

4. Hold the sensor at an angle and shake out the old

filling solution.

5. Using the dispenser cap on the filling solution

bottle, insert the tube into the bottom of the empty
reservoir as shown. Squeeze a steady stream of
solution into the reservoir until it overflows and no
bubbles are observed. Continue to add solution
while pulling the tube out of the reservoir.

6. Attach the reference junction to the sensor and

hand-tighten until firmly attached. Some filling
solution will overflow. Wipe the excess off the
sensor body.

7. Soak the sensor in tap water for at least 15 minutes.
8. Calibrate the sensor.
Note: If necessary, thoroughly clean the sensor connector to

remove filling solution: Using a disposable pipette, fill the
connector with isopropyl alcohol (70% - 100%). Shake to
dry. Repeat 3 times. Dry overnight. When thoroughly dry,
calibrate the sensor.

Replacing the Junction
Replace the junction when the sensor fails to calibrate
with a reasonable slope and offset, even after you have
replaced the filling solution.

1. Unscrew the reference junction and discard.
2. Replace the filling solution and screw in a new

reference junction.

3. Soak for 15 minutes, then calibrate the sensor.
Note: Keep the junction damp at all times to avoid a lengthy

rewetting process.

Cleaning
Begin with the most gentle cleaning method and
continue to the other methods only if necessary. Do not
directly wipe the glass bulb.

To clean the pH sensor, gently rinse with cold water.
If further cleaning is required, consider the nature of
the debris.

To remove crystalline deposits:
• Clean the sensor with warm water and mild soap.
• Soak the sensor in 5% HCI solution for 10 to 30 minutes.
• If deposits persist, alternate soaking in 5% HCI and

5% NaOH solutions.

To remove oily or greasy residue:
• Clean the sensor with warm water and mild soap.
• Methanol or isopropyl alcohol may be used for

short soaking periods, up to 1 hour.

• Do not soak the sensor in strong solvents, such

as chlorinated solvents, ethers, or ketones, such
as acetone.

To remove protein-like material, or slimy film:
• Clean the sensor with warm water and mild soap.
• Soak the sensor in 0.1M HCI solution for 10 minutes

and then rinse with deionized DI water.

After performing any of these cleaning methods, rinse
the sensor with water, then soak overnight in pH 4 buffer.

Sensor Removal
You can store the pH sensor on the instrument, or you
can remove the sensor. If you remove the sensor from
the instrument, seal the port with a port plug and place a
dust cap on the port side of the pH sensor.

Figure 3. Fill the pH reservoir

Figure 2. pH sensor with reference junction removed

Advertising
This manual is related to the following products: