3 2-point calibration, 100% and 0% saturation, Instruction sheet – In-Situ Calibration Kit for the optical RDO PRO Probe User Manual

Page 2

Advertising
background image

Instruction Sheet

2-Point Calibration

100% and 0% Saturation

1. Set up the calibration procedure as previously

described, and perform a water-saturated air
calibration.

2. Remove the water-saturated sponge from the

calibration chamber and fill the chamber to the fill line
with approximately 60 mL of fresh sodium sulfite
solution.

1

Fill line

2

Temperature sensor

2. Place the instrument into the solution. Leave at least

13 mm (0.5 in.) between the surface of the sensing
material and the bottom of the chamber.

3. Ensure that the temperature sensor is completely

submerged in the solution.

4. Allow at least 5 minutes for the temperature to

stabilize prior to performing the calibration procedure.

5. Once calibration is complete, remove the sensor, and

thoroughly rinse to remove all of the sodium sulfite.