21

calibrated.

ASCII SO<CR> Opens the dark shutter which allows monochromatic light to enter the

sample chamber. This allows 100% transmittance (zero absorbance) to
be calibrated. The shutter must be open for normal measurements.

ASCII Gnnn<CR> Commands the instrument to go to wavelength nnm. For example:

G540<CR> will set the wavelength to 540nm.

ASCII Fxxxx.x<CR> Sets the concentration factor to xxxx.x. For example: F1000<CR> will set

the factor to 1000.

Note <CR> is an ASCII carriage return character. The last three commands provide an output

which can be readily incorporated into most spreadsheet software packages.

6.1

RS232 Output

The bi-directional RS232 interface is available on the rear panel 25 way D type connector. The
connections are as follows:

TDX 2

– INPUT TO 6305

RXD 3

– OUTPUT FROM 6305

RTS 4

– LINKED TO CTS

CTS 5

– LINKED TO RTS

DSR 6

– OUTPUT FROM 6305

DCD 8

– OUTPUT FROM 6305

DTR 20

– INPUT TO 6305 (must be active)

GND 7

Suggested interconnections are detailed below:

6305

IBM PC XT (25 way “D”)

TDX 2

2 TXD (From PC)

RXD 3

3 RXD (To PC)

RTS 4

4 RTS (From PC)

CTS 5

5 CTS (To PC)

DSR 6

6 DSR (To PC)

DCD 8

8 DCD (To PC)

DTR 20

20 DTR (From PC)

GND 7

7 GND