SECTION 9 – Troubleshooting

9.1 ERROR CODES

Refer to the table below to diagnose error codes. If you require further assistance contact Jenway
Help (refer to section 9.2).

Error Code

Error Name

Issue

Err 1

Dark cal error


This error occurs when the 6305 closes the shutter to block

light entering the sample chamber. If the detector output

does not fall to a level normally associated with a dark cal

then this error is indicated. The most likely cause is that the

sample chamber lid is not closed.

Err 2

Light cal error




This error indicates that there is insufficient light to calibrate

to 100%. The most likely cause is that the light at the

selected wavelength is being absorbed by a sample in the

sample chamber.

This error can also be caused if the lamp has failed.

Err 3

Concentration
calibration error

The calculated factor is outside the instrument range.

Recalibrate the instrument using a standard of lower

concentration.

Err 5

Lamp Failure

The instrument is unable to detect a peak signal for

wavelength calibration. Possible causes of this error are that

the lamp has failed or that a sample has been left in the

sample chamber during start-up.

Err 6

Power on self-test
error

Detector signal with the lamp off exceeds the preset

threshold. Possible cause is that the sample chamber lid is

not closed.

Err 7

Power on self-test
error

The instrument is unable to detect the monochromator

reference position sensor. Please contact Bibby Scientific

service or your local distributor.

9.2

TECHNICAL SUPPORT

Jenway have a dedicated Technical Support team made up of experienced scientists who are on
hand to help with any applications advice and questions you may have about our products

and

how to use them. If you require any technical or application assistance please contact the team
at:

E-mail:

[email protected]

Phone: +44 (0)1785 810433

Advertising