SERIES EL 071414
B-7
B.30
[SOURce:]CONDuctance:TRANsient COMMAND
COND:TRAN
Syntax:
Short Form: [SOUR:]COND:TRAN <value>,<time>,<period>,<count>
Long Form: [SOURce:]CONDuctance:TRANsient <value>,<time>,<period>,<count>
Where:
<value> = <NR2> = the maximum conductance value of the transient (Siemans)
<time> = <NR2> = time (Seconds) from start to end of transient (0.0005 Second minimum)
<count> = <NR1> = number of times transient is to be repeated (default = 1, maximum = 65,000)
<period> (optional) = <NR2> = total time (Seconds) of repeating pulse train (must be at least 0.0005
greater than transient <time>)
Description: Configures a transient conductance pulse that is initiated by SYST:MODE:TRAN (see PAR.
B.100). See PAR. B.106 for details on using both Transient and Ramp commands at the same time.
Figure B-1 on page B-18 shows transient and ramp timing.
B.31
[SOURce:]CONDuctance:TRANsient? QUERY
COND:TRAN?
Syntax:
Short Form: [SOUR:]COND:TRAN?
Long Form: [SOURce:]CONDuctance:TRANsient?
Returns: <value>,<time>,<period>,<count> (see PAR. B.30 for definitions).
Description: Indicates Conductance Transient parameters established by COND:TRAN command (PAR.
B.32
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude] COMMAND
CURR
Syntax:
Short Form: [SOUR:]CURR[:LEV][:IMM][:AMP] <NR2>
Long Form: [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude] <NR2>
Where <NR2> = current setpoint
Description: Current Setting. The current setpoint affects the load only when the load is in Current mode of opera-
tion. Current setpoint is stored in non-volatile memory; value is restored upon power-up.
B.33
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude]? QUERY
CURR?
Syntax:
Short Form: [SOUR:]CURR[:LEV][:IMM][:AMP]?
Long Form: [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude]?
Returns <NR2> = current setpoint
Description: Indicates Current Setpoint.
B.34
[SOURce:]CURRent:PROTection[:LEVel] COMMAND
CURR:PROT
Syntax:
Short Form: [SOUR:]CURR:PROT[:LEV] <NR2>
Long Form: [SOURce:]CURRent:PROTection[:LEVel] <NR2>
where: <NR2> = current protection setpoint
Description: Establishes current protection setpoint. Current protection setpoint is stored in non-volatile mem-
ory; value is restored upon power-up. This is a user-defined value which, if exceeded during opera-
tion, causes a FAULT condition, forcing the load to disengage.
B.35
[SOURce:]CURRent:PROTection[:LEVel]? QUERY
CURR:PROT?
Syntax:
Short Form: [SOUR:]CURR:PROT[:LEV]?
Long Form: [SOURce:]CURRent:PROTection[:LEVel]?
Returns: <NR2> = current protection setpoint
Description: Indicates current protection setpoint.