B.36 [source:]current:protection:state command, B.37 [source:]current:protection:state? query, B.38 [source:]current:transient command – KEPCO EL Series Electronic Load Operator Manual P/N 243-1295 Firmware Version 4.00 to 4.xx User Manual
Page 76: B.39 [source:]current:transient? query, B.40 [source:]mode command, B.41 [source:]mode? query, B.36, Source, B.37, B.38

B-8
SERIES EL 071414
B.36
[SOURce:]CURRent:PROTection:STATe COMMAND
CURR:PROT:STAT 0
Syntax:
Short Form: [SOUR:]CURR:PROT:STAT 0
Long Form: [SOURce:]CURRent:PROTection:STATe 0
Description: Clears current protection level error.
B.37
[SOURce:]CURRent:PROTection:STATe? QUERY
CURR:PROT:STAT?
Syntax:
Short Form: [SOUR:]CURR:PROT:STAT?
Long Form: [SOURce:]CURRent:PROTection:STATe?
Returns: <bool> = 1 (current protection error occurred)
0 (current protection error has not occurred)
Description: Indicates whether a current protection error has occurred (1) or not (0).
B.38
[SOURce:]CURRent:TRANsient COMMAND
CURR:TRAN
Syntax:
Short Form: [SOUR:]CURR:TRAN <value>,<time>,<period>,<count>
Long Form: [SOURce:]CURRent:TRANsient <value>,<time>,<period>,<count>
Where:
<value> = <NR2> = the maximum current value of the transient (Amperes)
<time> = <NR2> = time (Seconds) from start to end of transient (0.0005 Second minimum)
<count> = <NR1> = number of times transient is to be repeated (default = 1, maximum = 65,000)
<period> (optional) = <NR2> = total time (Seconds) of repeating pulse train (must be at least 0.0005
greater than transient <time>)
Description: Configures a transient current pulse that is initiated by SYST:MODE:TRAN (see PAR. B.100).
See PAR. B.106 for details on using both Transient and Ramp commands at the same time. Figure B-
1 on page B-18 shows transient and ramp timing.
B.39
[SOURce:]CURRent:TRANsient? QUERY
CURR:TRAN?
Syntax:
Short Form: [SOUR:]CURR:TRAN?
Long Form: [SOURce:]CURRent:TRANsient?
Returns: <value>,<time>,<period>,<count> (See PAR. B.38 for definitions.)
Description: Returns the transient current pulse parameters established by CURR:TRAN (see PAR. B.38).
B.40
[SOURce:]MODE COMMAND
MODE
Syntax:
Short Form: [SOUR:]MODE <txt>
Long Form: [SOURce:]MODE <txt>
where: <txt> = CURRent, POWer, VOLTage, RESistance, CONDuctance, SHORT or OFF.
Description: Sets operating mode: When the operating mode is changed, the input is set to off. Send INPut ON to
apply the new mode setting to the load.
B.41
[SOURce:]MODE? QUERY
MODE?
Syntax:
Short Form: [SOUR:]MODE?
Long Form: [SOURce:]MODE?
Returns: <txt> = CURR, POW, VOLT, RES, COND, SHORT or OFF.
Description: Indicates operating mode.
B.42
[SOURce:]POWer[:LEVel][:IMMediate][:AMPlitude] COMMAND
POW
Syntax:
Short Form: [SOUR:]POW[:LEV][:IMM][:AMP] <NR2>
Long Form: [SOURce:]POWer[:LEVel][:IMMediate][:AMPlitude] <NR2>
Where <NR2> = power setpoint
Description: Establishes power setpoint. The power setpoint only affects the load when the load is in Power
mode of operation.