Technical data, 4 measurement accuracy – KROHNE WATERFLUX 3000 EN User Manual
Page 33

TECHNICAL DATA
6
33
WATERFLUX 3000
www.krohne.com
06/2013 - 400162663 - HB WATERFLUX 3000 R04 en
6.4 Measurement accuracy
Each flowmeter is standard wet calibrated under reference conditions by direct volume
comparison. The performance of the flowmeter is defined and documented in an individual water
meter calibration certificate.
Reference conditions
• Medium: water
• Temperature: +10...30°C / +50...86°F
• Pressure: 1 bar / 14.5 psi
• Inlet section: ≥ 3 DN
• Outlet section: ≥ 1 DN
• Electrical conductivity: ≥ 300 μS/cm
Figure 6-2: Flow velocity vs. accuracy
X
XX
X [m/s]: flow velocity
Y
YY
Y [%]: deviation from the actual measured value
Sensor diameter
Converter type
Accuracy
Curve
DN25...600 / 1"...24"
IFC 050
0.5% of mv above 0,5 m/s
below 0,5 m/s, deviation ± 2.5 mm/s
5
Sensor diameter
Converter type
Accuracy
Curve
DN350...600 / 14"...24"
IFC 100
0.5% of mv +1 mm/s
4
DN25...300 / 1"...12"
IFC 100
0.3% of mv +0.5 mm/s
2
DN350...600 / 14"...24"
IFC 300 + IFC 070
0.4% of mv +1 mm/s
3
DN25...300 / 1"...12"
IFC 300 + IFC 070
0.2% of mv +0.5 mm/s
1