Operation – KROHNE USB ADAPTER PLUS EMF EN User Manual
Page 8

2
OPERATION
8
USB ADAPTER PLUS
www.krohne.com
06/2014 - 4003748601 - QS USB_Adapter_Plus R03 en
Menu C1: Process input
No.
Function / subfunction
Settings / descriptions
C1.1 Calibration
C1.1
Calibration
C1.1.1
Zero calibration
Measured zero point (m/s).
Use the "calibration" button to calculate a new zero. Calibration procedure
takes 30 seconds.
C1.1.2
Size entry selection
Value (mm), Table (mm, inch)
For "Value": manual input of flow meter size in box 1.1.3.
For "Table": choose from pull down menu in box 1.1.4.
C1.1.5
GK selection
GK, GKL
For "GK": value is shown in box 1.1.6.
For "GKL": value is shown in box 1.1.7
C1.1.8
Coil resistance
10...220 Ω
Configured, can be changed.
C1.1.9
Field frequency
1/6...2 * line frequency
Choose from pull down menu.
1.1.10
Select settling
Standard (predefined), Manual
1.1.11
Settling time
0,001...0,25 (s)
Manual input
1.1.12
Line frequency
50 or 60 Hz
1.1.13
Measurement conductivity Off, On
Off: Default
1.1.14
Number values
0...10
Number of measurements taken to calculate the average conductivity (if
1.1.13 is switched on).