Accuracy (similar to DIN EN 60770-1)
Reference conditions acc. to DIN EN 61298-1
- Temperature
18 … +30°C (64 … +86°F)
- Relative humidity
45 … 75 %
- Atmospheric pressure
860 … 1060 mbar (86 … 106 kPa/
12.5 … 15.4 psi)
Characteristic curve deviation and measurement characteristics
3)
Average temperature coefficient of the
zero signal (temperature error)
0.06 %/10 K
Resolution, general
max. 1 mm
Ultrasonic frequency
55 kHz
Interval
> 2 s (dependent on the parameter adjustment)
Beam angle at -3 dB
5.5°
Adjustment time
4)
>3 s (dependent on the parameter adjustment)
Accuracy
better than 0.2 % or ±4 mm (see diagram)
8 m
4 m
3 m
2 m
7 m
6 m
5 m
1 m
10 mm
16 mm
4 mm
-4 mm
-10 mm
-16 mm
Fig. 22: Accuracy OPTISOUND 3020 C
3)
Relating to the nominal range, incl. hysteresis and repeatability, determined
acc. to the limit point method.
4)
Time to output the correct level (with max. 10 % deviation) after a sudden
level change.
38
OPTISOUND 3020 C - 4 ... 20 mA/HART - four-wire